Light incubator operating procedures

1 Purpose To regulate the operation and maintenance of the light incubator to ensure the normal operation of the instrument.

2 The scope of application is applicable to the operation and maintenance management of the light incubator in the plant tissue culture experimental training room.

3 Responsible person in charge of the experimental training room.

4 operating procedures

4.1 Adjust the height with the adjusting screw at the bottom of the box to make the box stable.

4.2 Plug in the power socket The power supply should have good grounding. Press the power switch to display the display. The display shows the actual temperature inside the incubator.

4.3 Temperature adjustment Press the temperature setting button. The digital display is the set value. Rotate the temperature adjustment potentiometer to the desired temperature value. The release button digital display is the actual temperature in the culture chamber. At this time, if the actual temperature in the incubator is smaller than the set temperature, the heating indicator lights up. The heater starts to heat. The actual temperature in the incubator is larger than the set temperature. The cooling indicator lights up. The cooling system starts to cool. If the heating indicator and the cooling indicator are both dark. The incubator is at a constant temperature.

4.4 The intensity of the illuminance is controlled by the six corresponding power fluorescent lamps on the left and right sides of the box. You can select the six switches on the panel to adjust the light intensity as needed.

4.5 Day and night running time control You can set day and night working parameters as needed. For example, if you want to set the daytime running time to 10 hours, the night running time is 14 hours. The specific operation is as follows

(1) Press the SET/RUN button to display “b××” and press the “▲” or “▼” button to display “b10”.

(2) Press the SET/RUN button again to display “h××” and press the “▲” or “▼” button to display “h14”.

(3) After setting the parameters, press the Set/Run button to enter the daytime running state. The daytime indicator light is on and the night light is off. When the daytime work time is over, the night work time is running. At this time, the night light is on. This cycle.

(4) Press the night start button in any state to enter the running night working time state.

(5) Entering the night running state, all the fluorescent lamps are automatically powered off.

4.6 When the temperature is set, the temperature control knob cannot be rotated back and forth many times to avoid the compressor starting frequently causing the compressor to overload and affect the service life of the compressor.

4.7 Care must be taken when handling and the angle between the water level should not be less than 45 °.

4.8 When the temperature is low, the water tank may be slightly inclined backwards.

4.9 If the fluorescent tube needs to be replaced, the left and right side light holders can be removed by removing the rear panel.

4.10 In order to keep the appearance of the equipment, it is not allowed to wipe the surface box with acid or alkali and other corrosive materials. It can be dried regularly with a dry cloth.

4.11 Unplug the power cord when the instrument is stopped.

Dehydrated Seaweed Tablets

The natural Organic Purple Sweet Potato in instant noodles is made of fresh seaweed, steamed to dryness at high temperature, and seasoned with salt, pepper, and chili. Natural Organic Purple Sweet Potato is marinated and soaked, then a piece of sun-dried seaweed is roasted. Cut into small pieces and it's Dehydrated Purple Sweet Potato.